Specifications
1.Patient type: Adult, pediatric and neonatal
2.Driven mode: Electrical motor driving
3.Applicable IV set:
One specified IV set
Four specifications of ordinary IV sets
4.Flow rate: 1mL/h~1300mL/h
5.Preset volume range: 1mL~9999mL
6.Accumulated infusion volume display range: 1mL~9999mL
7.KVO (keep vein open):
1mL/h ≤flow rate ≤5mL/h, KVO rate is OFF
5mL/h
50mL/h ≤flow rate <300mL/h, KVO rate is 3mL/h
300mL/h ≤flow rate ≤1300mL/h, KVO rate is 5mL/h
8.Flow rate accuracy:
±5% (specified IV set)
±10% after calibration (ordinary IV set)
PRODUCT DETAILS
Storage & Operation Condition
Storage:
Temperature: -20~55°c
Relative humidity: ≤93%
Atmospheric pressure: 50KPa~106KPa
No exposed to corrosive gases, stored in well ventilated room
Operation:
Temperature: 5~40°c
Relative humidity: 20~90%
Atmospheric pressure: 70KPa~106KPa
Altitude: 4000m below
Dimension & Weight
Dimension: H245xD180xW170mm
Weight: ≤3kg
Accessories & Document
Power calbe
User Manual
Baghdad Medical Devices Office