+964 790 142 6499 / +964 771 458 1494
Inner Highlight

Ditermo MB400

Electrosurgical equipments monopolar,bipolar cut and coagulation

DIATERMO MB400D are high frequency electrosurgical equipments suitable to deliver current for monopolar cut and coagulated cut (with different levels of coagulation), bipolar cut and coagulation.
 TWIN OPERATIVE ACTIVITY - two output channels for monopolar cutting - two independent setting and output channels for the most frequent monopolar coagulation
 CONTROLLED CUTTING PROCEDURES - enhanced cutting current for eschar free incision - cutting current with selectable coagulating property for bleeding controlled dissection - low capacitive coupling by slow BLEND current for endoscopic procedures
TWIN CONTROLLED COAGULATION PROCEDURES - SPEEDY coagulation and DEEP coagulation available to two operators - SPRAY coagulation by hand switch or foot switch controls
 EXTENSIVE BIPOLAR FEATURES
 - bipolar cutting current with selectable bleeding control - slow BLEND cutting current,
SURGEON AND PATIENT TOTAL SAFETY - acceptability of contact impedance between patient body and return electrode continuously monitored by Skin Plate Electronic Control (SPEC) - H.F. Leakage currents continuously monitored - mains fl uctuation automatically compensated before delivery of HF power PERSONALIZED PROCEDURES - independent choice of coagulation currents and power levels by each of two operators - up to 10 procedures can be confi gured and re-called
 EXTENSION DRIVER AND INTERLOCK SIGNALS - driven output connection to external Argon Plasma unit and Smoke Evacuator unit
 AUTOMATIC SELF DIAGNOSIS - automatic diagnostic control of the complete functionality of electronic unit and connected accessories with coded report of failures
Maximum out power monopolar pure CUT 400 W - 300 Ω
Maximum out power monopolar cut-coag CUT80% 350 W - 300 Ω
Maximum out power monopolar cut-coag CUT60 250 W - 300 Ω
Maximum out power monopolar cut-coag CUT40% 200 W - 300 Ω
 Maximum out power monopolar cut ENHANCED 250 W - 500 Ω
 Maximum out power monopolar cut-coag BLEND 250 W - 300 Ω
Maximum out power monopolar coag SPEEDY 120 W - 500 Ω
Maximum out power monopolar coag DEEP 120 W - 200 Ω
 Maximum out power monopolar coag SPRAY 100 W - 2000 Ω
 Maximum out power monopolar coag SPRAY ARGON 100 W - 2000 Ω
 Maximum out power bipolar CUT 95 W - 150 Ω
Maximum out power bipolar cut-coag CUT80% 95 W - 150 Ω
 Maximum out power bipolar cut-coag CUT60% 95 W - 150 Ω
 Maximum out power bipolar cut-coag CUT40% 60 W - 150 Ω
Maximum out power bipolar cut-coag BLEND 90 W - 150 Ω
Maximum out power bipolar cut-coag BLEND 70 W - 100 Ω
Working frequency monopolar / bipolar 425 kHz / 525 kHz
 Patient circuit F
 Selectable input voltage 115-230 Vac
 Mains frequency 50-60 Hz
 Electrical input voltage 1000 VA
 Size WxHxD mm 470x150x400
Weight 17.5 Kgs
Controls: Patient/Plate circuit monitoring, Output power monitoring, Self check control Safety: EN60601-1, EN60601-1-2, EN60601-2-2, Electrical Class: I CF, MDD 93/42/EC Class: II b

Baghdad Medical Devices Office

Baghdad - Saadoun Street - Al-Naser Square

info@baghdad-medical.com

+964 790 142 6499 / +964 771 458 1494