DIATERMO MB400D are high frequency electrosurgical equipments suitable to deliver current for monopolar cut and coagulated cut (with different levels of coagulation), bipolar cut and coagulation.
TWIN OPERATIVE ACTIVITY - two output channels for monopolar cutting - two independent setting and output channels for the most frequent monopolar coagulation
CONTROLLED CUTTING PROCEDURES - enhanced cutting current for eschar free incision - cutting current with selectable coagulating property for bleeding controlled dissection - low capacitive coupling by slow BLEND current for endoscopic procedures
TWIN CONTROLLED COAGULATION PROCEDURES - SPEEDY coagulation and DEEP coagulation available to two operators - SPRAY coagulation by hand switch or foot switch controls
EXTENSIVE BIPOLAR FEATURES
- bipolar cutting current with selectable bleeding control - slow BLEND cutting current,
SURGEON AND PATIENT TOTAL SAFETY - acceptability of contact impedance between patient body and return electrode continuously monitored by Skin Plate Electronic Control (SPEC) - H.F. Leakage currents continuously monitored - mains fl uctuation automatically compensated before delivery of HF power PERSONALIZED PROCEDURES - independent choice of coagulation currents and power levels by each of two operators - up to 10 procedures can be confi gured and re-called
EXTENSION DRIVER AND INTERLOCK SIGNALS - driven output connection to external Argon Plasma unit and Smoke Evacuator unit
AUTOMATIC SELF DIAGNOSIS - automatic diagnostic control of the complete functionality of electronic unit and connected accessories with coded report of failures
Maximum out power monopolar pure CUT 400 W - 300 Ω
Maximum out power monopolar cut-coag CUT80% 350 W - 300 Ω
Maximum out power monopolar cut-coag CUT60 250 W - 300 Ω
Maximum out power monopolar cut-coag CUT40% 200 W - 300 Ω
Maximum out power monopolar cut ENHANCED 250 W - 500 Ω
Maximum out power monopolar cut-coag BLEND 250 W - 300 Ω
Maximum out power monopolar coag SPEEDY 120 W - 500 Ω
Maximum out power monopolar coag DEEP 120 W - 200 Ω
Maximum out power monopolar coag SPRAY 100 W - 2000 Ω
Maximum out power monopolar coag SPRAY ARGON 100 W - 2000 Ω
Maximum out power bipolar CUT 95 W - 150 Ω
Maximum out power bipolar cut-coag CUT80% 95 W - 150 Ω
Maximum out power bipolar cut-coag CUT60% 95 W - 150 Ω
Maximum out power bipolar cut-coag CUT40% 60 W - 150 Ω
Maximum out power bipolar cut-coag BLEND 90 W - 150 Ω
Maximum out power bipolar cut-coag BLEND 70 W - 100 Ω
Working frequency monopolar / bipolar 425 kHz / 525 kHz
Patient circuit F
Selectable input voltage 115-230 Vac
Mains frequency 50-60 Hz
Electrical input voltage 1000 VA
Size WxHxD mm 470x150x400
Weight 17.5 Kgs
Controls: Patient/Plate circuit monitoring, Output power monitoring, Self check control Safety: EN60601-1, EN60601-1-2, EN60601-2-2, Electrical Class: I CF, MDD 93/42/EC Class: II b